|
|
Results:
Figure 9 - Frequency Drift of TDO measured at room temp.
Figure 10 - Frequency Drift of TDO measured at 77 Kelvin
Figure 11 - Frequency Drift of TDO measured at 4.5 Kelvin
One of the most pressing sources of systematic error in this experiment is due to frequency drift. Since measuring the susceptibility requires two different measurements; one with the sample in the inductor and one measuring the background frequency. Any drift in frequency during this period will lead to significant error. We measured the frequency drift at several different temperatures, specifically at room temperature (figure 9), at 77 K(figure10) and at 4.5 Kelvin (figure 11). One can easily see that the room temperature drift is quite large, varying by 900 Hz over a 20 minute data run. However when the circuit is cooled to 77K the frequency only drifts by 300 Hz during a 15 minute data run. For this data set the circuit was still cooling down in the first few data points, therefore they do not contribute to our frequency drift measurements. The drift at 4.5 Kelvin is even more impressive, over a 30 minute data run the frequency drifts only 20 Hz. However from the graph one can see anomalous 100 Hz dips in our data; we have been unable to explain these dips in frequency. Since the dip always returns to the original frequency we do not believe that it is frequency drift, but rather some unexplained noise. Since we will be measuring changes much greater than the frequency drift at low temperatures, this effect will not inhibit our measurements.
Before taking measurements at low temperature we took a background frequency vs.
temperature measurement. This is shown in figure 12. In the region between 2 and 30 K the
frequency can be approximated by four different functions, to find these functions we used
LSQ Fit, a least squares fitting program. Between 2 and 5.4 K the background frequency can
be approximated by f(T) = 31081459(±16.17) - 352(±4.19)T with a reduced
A. Measurements of Bismuth We initially attempted to measure the magnetic susceptibility of Bismuth at room temperature. For this experiment we used 99.5% to 100% purity Bismuth samples manufactured by Merck & Co. The samples were massed before being placed within the inductor coil. Their density at 25 C is 9.79g/cm3 (8).
Four different samples were measured, and our results are tabulated in Table 1. The
values have such high error because the frequency drift in this region was quite high
(Figure 9) Furthermore, as shown in Table 1, the calculated values of
B. Measurements of CrO8S2 12H2O
In order to measure the magnetic susceptibility of CrO8S2 12H2O we placed 0.0163 grams of the substance within a 0.11±0.005 cm inner diameter, 0.221±0.001 cm length of Teflon tubing; this equals 0.0020 ±0.0002 cm3, which corresponds to a filling factor of 0.046±0.009. The ends were capped with five minute epoxy, and the entire tubing was placed within the core of the inductor. The core was held snugly in place using 0.0138± 0.0001 grams of Teflon tape. The measurements were all made within the Quantum Devices Physical Properties Measurement System. We took two frequency vs. temperature measurements, one from 2-11.3 K, and the other from 13-30K. A graph of these measurements is shown in figure 13. After subtracting the background frequency data we can easily see that our data is not what we expected (Figure 14). As one can see there is actually an increase in frequency after putting the sample into the inductor. This is opposite the effect that would indicate paramagnetism. Furthermore as the temperature increases the frequency also increases. After calculating our magnetic susceptibility vs. temperature one can easily see that the data does not show the expected xm = C/T. We have yet to determine the cause of these readings. C. Measurements of Niobium
In order to measure the transition the transition temperature of Niobium we placed
0.0083±0.0001g of 0.14±0.005 mm diameter Niobium wire into the core of the
inductor. We then measured the frequency and amplitude vs. temperature from 4.9K to 14K.
The frequency vs. temperature data is shown in figure 18, and the amplitude vs. temperature
data is shown in figure 19. From this data one can easily see that the transition
temperature for this sample of Niobium starts at 8.2±0.1K and ends at 8.6±.1K. After
subtracting the background frequency from our data one can see that the frequency is
constant above the critical temperature, and below the critical temperature the frequency
is slightly increasing. Therefore since the susceptibility must be constant once the
critical temperature has been reached, the only changing property of the material is the
penetration depth. Therefore plotting |